ErrorDescriptionSolution
PAD: Abnormally weak pulse detected. Possibility of severe PAD or incompressible arteries. Please check the cuff position and repeat the measurement.The pulse was too weak. Reposition and repeat the measurement.Check the position of the cuff and repeat the measurement. Follow Chapter 5.3.1 Cuff placement.
E2: An anomaly has been detected. The patient may have moved during the measurement process.The pulse was too weak. Reposition and repeat the measurement.Remind the patient to remain still during measurement and repeat the measurement.
E3: Inflation insufficiency. Check the placement of the cuffs and repeat the measurement.Inflation insufficiency.Check the placement of the cuffs and repeat the measurement.
E4: An error occurred during heart rate calculation.An error occurred during heart rate calculation.Repeat the measurement. If the error persists, the measured value is outside the measurement range of the device.
E5: An error occurred during ankle brachial pressure index calculation.An error occurred during ankle brachial calculation.Repeat the measurement. If the error persists, the measured value is outside the measurement range of the device.
E6: Inflation error. Check the placement of the cuffs and repeat the measurement.An error occurred during inflation of the cuff.Check the placement of the cuffs and repeat the measurement.
E7: Deflation too fast. The cuff is not attached or is not attached correctly. Place the cuff correctly and repeat the measurement.The cuff was insufficiently deflated.The cuff is not attached or is not attached correctly. Attach the cuff correctly and repeat the measurement.
E8: An error occurred during systolic pressure calculation.An error occurred during systolic pressure calculation.Repeat the measurement. if the error is repeated, the measured value is outside the measurement range of the device.
E9: An error occurred during diastolic pressure calculation.An error occurred during diastolic pressure calculation.Repeat the measurement. if the error is repeated, the measured value is outside the measurement range of the device.
E10: An error occurred during mean pressure calculation.An error occurred during mean pressure calculation.Repeat the measurement. if the error is repeated, the measured value is outside the measurement range of the device.
E11: A large pressure fluctuation was detected. The patient may have moved.A large pressure fluctuation was detected. The patient may have moved.The result may be incorrect. Repeat the measurement.